Troubleshooting
Error displayed
Definition
Attention!
Readings with barcode
cuvettes generally
Please insert the light
require the light shield.
shield.
Barcode label not read
Barcode faulty
An error occurred
when uploading the
instrument data.
An error occurred when
reading from the USB
memory stick.
An error occurred when
writing to the USB
memory stick.
Please check on the
Error during update.
current update file.
Please contact
Error during update.
Customer Services.
Please check network
configuration.
Please check the
connection.
Please close the cover.
English 118
Error displayed
Please insert the USB
Resolution
memory stick.
Insert the light shield.
Please check the
connection and contact
Press OK.
the administrator.
Reinsert the cell.
Blank value correction
If the barcode is not
not possible!
recognised, contact the
manufacturer or his
File for instrument
representative.
update missing.
Start the procedure again
or contact the
File for instrument
manufacturer or his
update is faulty.
representative.
Start the procedure again
or contact the
It's recommended to
manufacturer or his
execute a Full System
representative.
Check
Start the procedure again
or contact the
manufacturer or his
representative.
Entry invalid!
Check the USB memory
stick.
Absorbance > 3.5!
Fault
Barcode check
number?
Please update program
data!
Close the cover.
Error when calling up
the local IP address.
Definition
Resolution
Insert a USB memory
stick into a USB A port on
the instrument.
Network setup or FTP
error
Blank value correction
not possible with
LCW919.
Check the USB memory
Error during update.
stick.
Save the update file
Error during update.
again and repeat the
procedure.
Switch the instrument of
and then back on again.
Check of the air values
If the system check is not
failed
successful, contact the
manufacturer or his
representative.
Forgotten your
password?
Password incorrect
Contact the
manufacturer or his
representative.
The measured absor-
Dilute sample and
bance exceeds 3.5
measure again
Deviation to the stored
Data updating
data
Network setup: DHCP cli-
Enter the IP address
ent has no connection to
again.
the DHCP server