Operating Environment - Hach DR 3900 Manual Básico Del Usuario

Ocultar thumbs Ver también para DR 3900:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 124
Installation
WA R N I N G
Electrical and fire hazards.
Only use the supplied benchtop power supply LZV844.
Only qualified experts may perform the tasks described in this section of the man-
ual, while adhering to all locally valid safety regulations.
Unpack the instrument
The DR 3900 Spectrophotometer comes packaged with the following
items:
DR 3900 spectrophotometer
Dust Cover
USB dust cover, fitted as standard
Benchtop power supply with power cord for EU
Cell adapter A
Light shield, fitted as standard in the DR 3900
Operator RFID tag (not available on all models)
Basic operating instructions DR 3900, operating instructions
LINK2SC
Further information, detailed user manuals and documentation are
available on the website of the manufacturer.
Note: If any of these items are missing or damaged, please contact the
manufacturer or a sales representative immediately.
English 106

Operating environment

Note the following points so that the instrument will operate perfectly and
have a long service life.
Place the instrument firmly on an even surface. Do not push any
objects under the instrument.
The ambient temperature must be 10–40 °C (50–104 °F).
The relative humidity should be less than 80 %; moisture should not
condense on the instrument.
Leave at least a 15 cm clearance at the top and on all sides for air
circulation, to avoid overheating of electrical parts.
Do not operate or store the instrument in extremely dusty, humid or
wet locations.
Keep the surface of the instrument, the cell compartment and all
accessories clean and dry at all times. Immediately remove
splashes or spilt materials on or in the instrument.
N O T I C E
Protect the instrument from extreme temperatures from heaters, direct sunlight
and other heat sources.

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido