Section 5 Accessories
Immersion
Cooler Bridge
Installation
CAUTION
5-2
Tools required:
• Phillips Head screwdriver
Procedure:
Always turn off the circulator and disconnect the power cord from
the power source before installing the bridge.
1.
Undo the four thumbscrews securing the immersion circulator to the
top panel and remove the immersion circulator.
2.
Undo the four Phillips Head screws securing the top panel to the bath
and remove it.
3.
Turn the old panel over and note the placement of its three gaskets.
Using the old panel as a template, install the three supplied gaskets in
the same position on the new panel. Note Place the panels on a soft
clean cloth, their stainless steel surfaces are susceptible to scratching.
4.
Place the immersion cooler bridge on the bath and secure it to the
circulator using the four Phillips Head screws.
5.
Place the immersion circulator on the top panel and secure it using the
four thumbscrews, hand tight.
6.
Remove the two screws securing the "dummy" panel to the immersion
cooler bridge.
7.
Insert the immersion circulator head through the hole.
8.
Secure the head to the top panel using the two supplied panels.
Immersion
Cooler Head
Thermo Scientific