Section 3 Installation
Additional Fluid
Precautions
DANGER
CAUTION
CAUTION
3-12
Ensure, when selecting the heat transfer fluid, that no toxic gases are
generated. Flammable gases can build up over the fluid during usage.
Ensure the over temperature cut-off point is set lower than the fire
point for the heat transfer fluid selected.
The highest working temperature as defined by the EN 61010-2-010 must
be limited to 25°C below the fire point of the bath fluid.
When working with fluids other than water:
• Do not use any fluid until you have read and understood the label and
the Material Safety Data Sheet (SDS).
• Do not blend any fluids.
• Ensure any fluid residue or any other material is thoroughly removed
before filling the bath with a different fluid.
• Always wear protective clothing, especially a face shield and gloves.
• Avoid spattering on the circulator's components, always slowly add fluid.
When adding fluid, point the opening of a container away from yourself.
• Use fume hoods.
• Do not allow any ignition sources in the vicinity.
Thermo Scientific