Ambient
Conditions
Immersion
Circulator Only
DANGER
CAUTION
CAUTION
Thermo Scientific
Installation
Section 3
Ambient Temperature Range
Maximum Relative Humidity
Operating Altitude
Overvoltage Category
Pollution Degree
Degree of Protection
*decreasing linearly to 50% relative humidity at 40°C (104°F)
Ensure the label and small piece of Styrofoam securing the float switch are
removed before using.
The circulator is designed for continuous operation and for indoor use.
For stand-alone immersion circulators use the supplied nut and washer
to secure the grounding strap to the top of the metal tank. See next
page.
Never place the circulator in a location where excessive heat, moisture,
inadequate ventilation, or corrosive materials are present.
Carefully install the circulator to ensure it does not fall into the bath or
that its line cord does not make contact with the bath contents.
For immersion circulators equipped with a bridge:
•
The bridge is designed to fit baths that are 400 mm to 800 mm wide.
•
Slide the bridge support rods to the desired length and secure them in
place with the supplied eight 3 mm Phillips Head screws.
•
If possible, secure the bridge to your bath using the two supplied thumb-
screws.
•
The immersion depth is 75 to 145 mm (~3 to 5
Your bath container must be sturdy enough to support the weight of
the assembly, approximately 3.8 kilograms (8.5 pounds). The circulator is
designed for continuous operation and for indoor use only.
5°C to 40°C (41°F to 104°F)
80% for temperatures up to 31°C (88°F)*
Sea Level to 2000 meters (6560 feet)
II
2
IP 20
3
/
").
4
3-1