Cavitron
300 Series
®
Ultrasonic Scaling System
Maintenance Guide
IMPORTANT: Turn OFF water supply to Cavitron system or disconnect water quick-disconnect from
the water supply when not in use.

DAILY MAINTENANCE

START-UP PROCEDURES AT THE
BEGINNING OF THE DAY:
1. Open the manual shut-off valve on the dental office
water supply system.
2. Install a sterilized Steri-Mate Handpiece onto the
handpiece cable.
3. Set the Power Level Control to minimum and the
Lavage Control to maximum.
4. Turn the system ON.
5. Hold the sterilized handpiece (without insert installed)
over a sink or drain. Activate the Purge Mode by
pressing the Purge icon.
• The Purge screen will appear indicating proper
activation of the purge function. Time icon will
count down indicating completion of purge cycle.
• The Purge function can be interrupted at any time
during the two minute cycle by pressing the Scale
icon or by pressing the Foot Pedal.
6. After completing the purge cycle, place a sterilized
30kHz Cavitron Ultrasonic Insert into the handpiece
and set the Power Level Control and Lavage Control to
your preferred operating position for ultrasonic scaling.
BETWEEN PATIENTS:
1. Remove the used Cavitron Ultrasonic Insert. Clean and
sterilize following the Infection Control Procedures that
were enclosed with your insert.
2. Hold the handpiece over a sink or drain and activate
Purge function as described in Step 5 of the
Start-Up procedure.
3. After the Purge cycle is complete, turn the System to
the OFF (0) position.
4. Remove the Steri-Mate handpiece, clean and sterilize
following the procedures outlined in the Cavitron Systems
Infection Control Procedures booklet that was enclosed
with your system.
5. Before disinfecting surface of cabinets, lock the touch
screen by pressing the Lock icon. Disinfect the surfaces
of the cabinet, Power Cord, Handpiece Cable, Tap-On
Foot Pedal and Cable assembly (if applicable), Water
Supply lines as instructed in the Cavitron Systems
Infection Control Procedures booklet.
DO NOT SPRAY DISINFECTANT SOLUTION DIRECTLY
ON SYSTEM SURFACES*.
Manufactured by:
Dentsply Professional
1301 Smile Way
York, PA 17404 USA
www.dentsplysirona.com
6. Inspect the handpiece cable for any breaks or tears
7. If using a closed water supply or DualSelect Dispensing
System, check for adequate fluid volumes for the next
patient.
8. When ready to use, press the unlock icon, place a sterilized
Steri-Mate Handpiece onto the handpiece cable assembly
and insert a sterilized ultrasonic insert into the handpiece
and adjust system controls as preferred.
SHUT-DOWN PROCEDURES AT THE
END OF THE DAY:
Follow the "Between Patients" maintenance procedures,
Steps 1 through 6. In addition, it is recommended to close the
manual shut-off valve on the dental water supply system.
*NOTE: Broad spectrum hospital grade water-based
disinfection solutions are preferred, such as Lysol IC. Some
alcohol-based disinfectant solutions may be harmful and may
discolor plastic materials.
WEEKLY MAINTENANCE
It is strongly recommended that this system be disinfected
by chemically flushing the waterlines with a 1:10 Sodium
Hypochlorite solution (NaOCl) at the end of each week.
This can be accomplished by connecting this device to the
Cavitron DualSelect Dispensing System or a number of other
devices available from your local distributors. Where this
device is connected to the Cavitron DualSelect Dispensing
System, please follow the DualSelect System's Directions for
Use manual. If connected to another device, please follow
those directions for use, keeping in mind that a chemical
flush should be performed at maximum water flow for at
least 30 seconds. The system should be left undisturbed for
10 minutes but no more than 30 minutes to allow the sodium
hypochlorite solution to soak in the lines. As a suggestion, it
is recommended that a sign be placed on the system stating
that the SYSTEM IS BEING DISINFECTED WITH A STRONG
DISINFECTANT AND SHOULD NOT BE USED. When ready,
flush system with clean water for at least 30 seconds or until
sodium hypochlorite odor disappears. ALL CHEMICALS MUST
BE FLUSHED FROM THE SYSTEM BEFORE IT IS READY FOR
PATIENT USE.