Calibrating Ultrasonic Height Measurement - Seca 655 Manual De Instrucciones

Ocultar thumbs Ver también para 655:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Calibrating ultrasonic height
measurement
3. Press the Filter item.
The current setting is displayed.
Filter
Low
Medium
High
4. Press the desired filter level.
The setting is active.
NOTE
With the Low setting, it may occur that no weight value is displayed
permanently despite the activated Hold function in the case of
patients who are not very steady on their feet.
Device mode
Basic
Advanced
Expert
Service
Before performing a measurement with the device for the first time, height
measurement must be calibrated. Repeat this calibration at least once per
year.
The automated calibration procedure consists of two steps:
• Calibration over the entire measuring range
• Calibration against a calibration rod (included in the scope of delivery of
the measuring rod).
NOTE
Make sure that there are no objects or people in the immediate vicinity
of the device during the calibration procedure ➔
device, page
93.
1. Press the
key.
The Settings menu is displayed.
2. Press the
or
displayed.
3. Press the Height meas. menu item.
Weight
determination
Medium
Function available
key until the Height meas. menu item is
English
Fast
Slow
Setting up the
Configuration • 109

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

654Mbca 555Mbca 554Mbca 552

Tabla de contenido