CARE & MAINTENANCE
Cuff Care & Maintenance
The cuff may be spot cleaned with a mild-detergent as needed.
To disinfect the cuff, wipe the inside (skin contact side) of the cuff using a cloth
lightly moistened with 70% - 90% ethyl rubbing alcohol and allow to air dry;
it is suggested to disinfect the cuff frequently when used in clinical settings or if
the cuff is contaminated in any way.
Never attempt to iron or dry the cuff; if the cuff becomes wet, allow the cuff to
air dry completely before use or storage.
Do not attempt to remove the tube and bladder from the nylon cuff as this may
damage the cuff.
Take care not to forcefully bend or kink the tubing during storage.
It is recommended the performance should be checked every 2 years or after
a repair. Please contact Customer Care at 866-326-1313.
It is recommended to clean the cuff every 200 uses.
The monitor can maintain the safety and performance characteristics for
minimum of 10,000 measurements of three years.
The cuff integrity is maintained after 1,000 open- close cycles of the closure.