Error Codes; Corrective Action - smartheart 01-539 Manual De Instrucciones

Automático monitor para la presión arterial
Ocultar thumbs Ver también para 01-539:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ERROR CODES

Code
E0
Pressure system is unstable before
measurement.
E1/ E2
Failed to detect systolic/ diastolic pressure.
E3/ E4
Pneumatic system leakage or cuff is too loose/
too tight during inflation.
E5
The cuff pressure is above 300mmHg.
More than 3 minutes with cuff pressure above
E6
15 mmHg.
EEPROM accessing error.
E7
Device parameter checking error.
E8
Pressure sensor parameter error.
ER A
Meaning

Corrective Action

Re-apply the cuff and attempt new
measurement.
Sit comfortable and remain still; ensure the cuff
is at the same level as your heart.
Re-apply the cuff and attempt new
measurement.
The measurement range is over 300 mmHg;
retake measurement and if error persists, consult
your physician.
Re-apply the cuff and attempt to take a new
measurement. Sit comfortable and remain still;
attempt new measurement.
ENGLISH • 27

Publicidad

Capítulos

Tabla de contenido

Solución de problemas

loading