Problem
VPAP Auto unit does
not stop when you
remove your mask.
SmartStop is enabled
but the flow
generator does not
stop automatically
when you remove
your mask.
Pressure rises
inappropriately.
Possible Cause
There is excessive leak.
SmartStop is disabled.
Note: SmartStop is not available
if you are using a full face mask
or if Leak Alert is enabled.
Incompatible humidifier or mask
system being used.
Talking, coughing, voluntarily
breath-holding (eg while rolling
over in bed), or intentionally
breathing in an unusual manner.
Mask cushion is buzzing against
the skin. Cushion seated
incorrectly.
Cushion seated incorrectly
causing excessive leak.
Air tubing is kinked or
punctured.
Humidifier control dial set too
high, resulting in accumulation
of water in the air tubing.
Solution
Adjust position of mask and
headgear.
Caps may be missing from ports
on mask. Replace them.
Air tubing not connected
properly. Connect firmly at both
ends.
Air tubing kinked or punctured.
Straighten or replace.
Enable SmartStop if possible.
Use only equipment as
recommended and supplied by
ResMed.
Avoid talking with a nasal mask
on, and breathe as normally as
possible.
Adjust the headgear.
Adjust headgear or re-fit
cushion.
Straighten or replace air tubing.
Turn humidifier control dial down
and empty the water from the
air tubing.
Troubleshooting
25