Problem
Display error
message:
Check tube!!
Key if done
Display error
message:
SYSTEM ERRnnnn
Call Service!
Display error
message: HIGH
LEAK!!!
Adjust Mask
Display error
message:
Exxxx (where xxxx
defines an error)
The following
message is displayed
on the LCD after you
try to update settings
using the Data Card:
Card Error
Remove Card
26
Possible Cause
The air tubing is loose or
blocked.
Component failure.
You have experienced
excessively high leak levels for
more than 20 seconds.
Component failure.
The Data Card is not inserted
correctly.
The Data Card is only partly
inserted.
You may have removed the Data
Card before settings were
copied to the VPAP Auto.
Solution
Check that the air tubing is
connected securely to your
mask and the air outlet on the
front of the VPAP Auto.
When you have checked the air
tubing, press the Start/Stop key
to restart the device.
If this does not clear the
message, disconnect the power
cord and then reconnect it to
restart the device.
Record error number and
contact your ResMed service
center.
Check that your air tubing is
connected properly and that
your mask does not leak
excessively.
Use the mask-fitting feature to
help you to fit your mask
properly.
If this message appears again,
contact your clinician.
Return your VPAP Auto for
servicing.
Ensure that the Data Card is
inserted with the arrow facing
up and pointing towards the
device. Ensure that the device is
switched on and in standby
mode before you insert the Data
Card.
Ensure that the Data Card is
inserted into the slot on the
Data Card module as far as it will
go.
Reinsert the Data Card and wait
for the Settings Success
Remove Card message to
appear on the LCD.