Electromagnetic compatibility (EMC) environ-
ment
This equipment is for use in home healthcare
facility environment.
This equipment is not likely to cause any
harmful interference in nearby equipment;
however, it can be impaired by electromagnetic
EN
(EM) interference emitted by other equipment,
which can lead to malfunctions. Do not use
this equipment where there is a risk that other
electrical and/or electronic equipment can
interfere with this equipment.
This equipment complies with the following
standards
IEC
60601-1-2:2014
equipment - Part 1-2: General requirements
for basic safety and essential performance
-
Collateral
disturbances - Requirements and tests (EN
60601-1-2:2015)
EN 12182:2012, Clause 7, Assistive products for
persons with disability – General requirements
and test methods
Testing for EM disturbances (EM emission and EM immunity)
Test modes during emission tests:
1.
Movement mode (battery). Actuator running in/out
2.
Charging mode, actuator not running.
EM emission standards
Conducted and radiated RF emission CISPR
11
Harmonic distortion
IEC 61000-3-2
Voltage fluctuations and flicker
IEC 61000-3-3
Medical
electrical
Standard:
Electromagnetic
EMC - WARNINGS
Use of this equipment adjacent to or stacked with
other equipment should be avoided because it
could result in improper operation. If such use
is necessary, this equipment and the other
equipment should be observed to verify that they
are operating normally.
Use of accessories, transducers and cables
other than those specified or provided by
R82 could result in increased EM emission or
decreased EM immunity of this equipment and
result in improper operation. Several power
plugs are supplied with this equipment. Please
use the correct power plug suitable for your
country. Please chose among the supplied power
plug adaptor units: English, Australian, EU, US/
Japanese plugs.
Portable radio frequency (RF) communication
equipment (including peripherals such as
antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any
part of the Caribou, including cables specified by
R82. Otherwise, degradation of the performance
of this equipment could result.
Classification/levels
Group 1 (the product uses RF energy only for its
internal function)
Class B (the product is suitable for use
in
all
establishments,
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes)
Class A.
No testing performed, as the rated power for
this equipment is less than 75 W. No limits are
specified for equipment with a rated power of 75
W or less.
dmax ≤ 4 % (without additional conditions)