Hose System; Heating Bath Hb 10 - VWR RV 8 Indicationes De Seguridad

Tabla de contenido

Publicidad

Also available with coating (RV10.40)
• RV10.5 Vertical-condenser with manifold and cut-off valve for reflux distillation
Also available with coating (RV10.50)
• RV10.6 Vertical-intensive condenser with manifold and cutoff valve for reflux distillation
This vertical-intensive action condenser features a double jacket design for particularly efficient reflux
distillation.
Also available with coating (RV10.60)

Hose system

Heating bath HB 10

Refer to the chapter entitled "Operation" in the heating bath instruction manual!
• Assemble the foot of the heating bath HB10
For operation of the HB10 heating bath in combination with the RV8, the foot plate supplied must be fitted to
the heating bath.
Caution: It is not permitted to use HB10 heating bath fitted with a foot plate in combination with any other type
of rotary evaporator.
• Connect the water hoses (H
according to the counter-flow principle.
• Install the vacuum connections to the condenser, Woulff
bottle, vacuum controller with valve and vacuum pump.
• Always connect the vacuum hose (vac) to the condenser at
the highest point to minimize solvent losses during
suctioning.
• Use standard laboratory vacuum hoses with an inside
diameter of 8 mm and a wall thickness of 5 mm (see User
Replaceable Accessories and Spare parts).
19
O) to the condenser
2

Publicidad

Tabla de contenido
loading

Tabla de contenido