1 Expanded manual version
2 Specifications
on page 3
3 General information
on page 4
4 Installation
on page 7
Section 1 Expanded manual version
For additional information, refer to the expanded version of this manual, which is available on the
manufacturer's website.
Section 2 Specifications
Specifications are subject to change without notice.
2.1 Inductive conductivity sensor
Sensitive measurement equipment. High levels of electrical interference can make measurements less accurate.
Keep electrical interference away from the power sources, probes and cables of measurement equipment.
Specification
Dimensions
Temperature element
Sensor cable
Wetted materials
Temperature/pressure limit Polypropylene: 100 °C at 6.9 bar (212 °F at 100 psi); PVDF: 120 °C at 6.9 bar
Operating temperature
Conductivity range
Temperature range
Maximum flow rate
Warranty
2.2 Inductive conductivity digital gateway
Specification
Dimensions (L x ∅)
Weight
Operating temperature
Humidity
Certifications
on page 3
N O T I C E
Details
Refer to
Figure 1
on page 6.
PT1000
5-conductor (plus two isolated shields), 6 m (20 ft); rated at 150 °C (302 °F)—
polypropylene
Polypropylene, PVDF, PEEK or PTFE
(248 °F at 100 psi); PEEK and PTFE: 200 °C at 13.8 bar (392 °F at 200 psi)
–10 to 200 °C (14 to 392 °F); limited only by sensor body material and mounting
hardware
0.0 to 200.0 µS/cm; 0 to 2,000,000 µS/cm
–10 to 200.0 °C (14 to 392 °F) limited by sensor body material
3 m/s (10 ft/s)
1 year; 2 years (EU)
Details
17.5 x 3.4 cm (7 x 1.4 in.)
145 g (5 oz)
–20 to 60 °C (–4 to 140 °F)
95% humidity, non-condensing
cETLus, CE, 3-A Sanitary Standard
5 User navigation
on page 15
6 Operation
on page 15
7 Maintenance
on page 22
8 Troubleshooting
on page 23
English 3