8019968/V1-0/2016-06 | SICK
Subject to change without notice
Press key "1" to go to the "meas. value outputs" menu.
6
settings 621
meas.value outputs
1 meas.value output 1
2 meas.value output 2
3 meas.value output 3
4 meas.value output 4
Enter digit
Back: ESCAPE
No messages
CO
12 mg/m3
The different measuring components are assigned to individual measured value outputs.
Perform the following steps successively for each measured value output and perform the
described changes as required.
Press key "1" to go to the "meas. value ouput 1" menu.The measuring component
7
assigned to measured value output 1 is shown (here: CO).
settings 6211
meas.value output 1
CO
1 measuring component
2 output range 1
3 output range 2
4 range selection
5 live zero (mA)
6 output range list
7 output assignment
8 delete config.
Enter digit
Back: ESCAPE
No messages
CO
12 mg/m3
Press key "2" to go to the "output range 1" menu.
8
settings 62112
output range 1
CO
1 beginning value
2 end value
3 switch-up value
Enter digit
Back: ESCAPE
No messages
CO
12 mg/m3
← measured value outputs
← output range 1
← end value
A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | SIDOR E
2
ENGLISH
15