L-3 conVersIon KIt adaPter InstaLLatIon InstrUctIons
1
1. Remove all existing nozzle parts. Nozzle
receptacle must be free of all obstructions
and/or loose parts before proceeding. Use 6mm
hex key provided to remove existing adapter.
2
3
L-3 conVersIon KIt User InstrUctIons
n
2. Verify new adapter
o
p
o-rings
and
in
position as shown.
3. Hand-thread adapter
into syringe. Use hex
key provided to tighten.
WarnIng: Do not over -
tighten! Damage to syringe
will result.
1
1. With thumb nut loosened, insert new Sani-Tip
push into adapter as far as possible. While still pushing inward,
twist tip 1/2 turn clockwise to create a watertight seal over internal
tapered water spindle.
2. Hand tighten thumb nut. To redirect tip, push in on tip while
2
rotating clockwise.
test PrIor to eacH Use: Express water, then express air only.
If air contains moisture, or tip dislodges, refer to Troubleshooting
section.
3. To replace tip, loosen thumb nut. Remove and discard used tip.
Repeat steps 1 and 2.
has
4
m
4. Verify o-ring
is in
place inside large
opening of white
l
plastic cone
.
5. With o-ring in place,
5
firmly snap large end
of white plastic cone
over adapter rim.
6. Hand-thread thumb
nut onto adapter. Do
6
not tighten until new
tip has been inserted.
Installation is complete.
disposable tip and
®
3