L-2 conVersIon KIt User InstrUctIons
InfectIon controL ProcedUres
WarnIngs
• Conversion Kits are not sterile upon receipt and must be sterilized prior to use in accordance with
the following instructions.
LIMItatIons on ProcessIng
• Do not use chemical disinfectants prior to sterilization or rapid deterioration of the material may
occur.
• Cold liquid disinfection/sterilization, chemical vapor sterilization, and dry heat sterilization methods
have not been tested or validated for efficacy and are not recommended for use. Use may void
warranty.
InstrUctIons
Point of use:
Remove excess soil with disposable cloth or paper.
containment and transportation:
Protect conversion kit components from damage. Components should be reprocessed as soon as is
reasonably practical following use. Components left standing wet may stain or corrode.
Preparation for decontamination:
Remove excess debris from the components using a disposable cloth or paper. Discard cloth/paper after
use. Disassemble the conversion kit from the air/water syringe. Remove the plastic cone and
adapter o-ring.
1
1. With thumb nut loosened, insert new Sani-Tip
push into adapter as far as possible. While still pushing inward,
twist tip 1/2 turn clockwise to create a watertight seal over internal
tapered water spindle.
2
2. Hand tighten thumb nut. To redirect tip, push in on tip while
rotating clockwise.
test PrIor to eacH Use: Express water, then express air only.
If air contains moisture, or tip dislodges, refer to Troubleshooting
section.
3. To replace tip, loosen thumb nut. Remove and discard used tip.
Repeat steps 1 and 2.
disposable tip and
®
5