5.2.2.12.Dipole vector automatic
• Push button [7] to select . Use the central controller [16] to link the channels;
• Push button [8], select
• Selection of automatic dipole vector current:
Select
with the central controller [16] and press button [14] to accept;
• Carrier wave selection:
Push button [9] to select
• Beat frequency:
Select
with button [10] and adjust its value with the central controller [16];
• Modulation frequency (spectrum):
Select
with button [10] and adjust its value with the central controller [16];
• Modulation time (spectrum):
Select
with button [11] and adjust its value with the central controller [16];
• Treatment time:
Select
with button [12] and adjust its value with the central controller [16];
• Rotation speed
Select
with button [12] and adjust its value with the central controller [16];
• Select CC or CV with button [13] and the central controller [16];
• Adjust the current intensity with controller [17].
5.2.3. Ultrasound
• Select
with button [7] and select
• Ultrasound frequency:
Select
(MHz) with button [9] and adjust its value with the central controller [16];
• Continuous or pulsed operation:
Select
with button [10] and adjust its value with the central controller [16]:
• Continuous = 100 %;
• Pulsed = 80%, 50%, 20%, 15%, 10% and 5 % with a frequency of 100Hz;
• Treatment time:
Select
with button [12] and adjust its value with the central controller [16];
• Select W or W/cm² with button [13] and the central controller [16];
• Adjust the ultrasound intensity with controller [18];
• Apply the ultrasound treatment head to the patient, at sufficient acoustical contact the applicator light will
be switched off and the treatment will be started.
5.2.4. Combination therapy
• Select
with button [7] and select
• Ultrasound frequency:
Select
(MHz) with button [9] and adjust its value with the central controller [16];
• Continuous or pulsed operation:
Select
with button [10] and adjust its value with the central controller [16]:
• Continuous = 100 %;
• Pulsed = 80%, 50%, 20%, 15%, 10% and 5 % with a frequency of 100Hz;
• Push button [8], select a current waveform with the central controller [16] and press button [14] to accept;
• Set all electrotherapy parameters according to paragraph 5.2.2;
• Treatment time:
Select
with button [12] and adjust its value with the central controller [16];
• Select W or W/cm² with button [13] and the central controller [16];
• Adjust the ultrasound intensity with controller [18];
• Apply the red electrode from channel 2 to the patient;
• Apply the ultrasound treatment head to the patient, at sufficient acoustical contact the applicator light will
be switched off and the treatment will be started;
• Adjust the current intensity on channel 2 with controller [17].
Attention: when combination therapy is selected, the apparatus will automatically select CV mode. To
change to CC mode, push button [13] and select CC with the central controller [16].
with the central controller [16] and press button [14] to accept;
(kHz) and adjust its value with the central controller [16];
with the central controller [16];
with the central controller [16];
41