ELECTRICAL SPECIFICATION

Ground Resistance (Auto range):
2
Range
Resolution
0.025 - 0.250 Ω
0.002 Ω
0.250 - 9.999 Ω
0.02 Ω
10.00 - 99.99 Ω
0.04 Ω
100.0 - 199.9 Ω
0.4 Ω
200.0 - 400.0 Ω
2 Ω
400.0 - 600.0 Ω
5 Ω
600.0 - 1550 Ω
20 Ω
1. Loop resistance non-inductive, external field <50 A/m, external electrical field <1 V/m, conductor
centered in jaws.
2. Resistance Measurement Frequency: 3.333 KHz
High and Low Alarm
Range
0 - 1550 Ω 1 Ω
High Alarm
0 - 1550 Ω 1 Ω
Lo Alarm
Ground/Leakage Current
(Auto range, 50/60 Hz, True RMS, Crest Factor <3.5)
Range
Resolution
0.200 - 1.000 mA
0.001 mA
1.00 - 10.00 mA
0.01 mA
10.0 - 100.0 mA
0.1 mA
100 - 1000 mA
1 mA
Ground/Leakage Current (50/60 Hz, True RMS, Crest Factor < 3.5)
Range
Resolution
0.20 – 35.00 A
0.01 A
Accuracy of Resistance Calibration Plate: ±1%
Data Logging Capacity: 8180 records (DET20C), 116 records (DET10C)
Data Logging Interval: 1 to 255 seconds
RS-232C Output: 0.5 and 1 to 255 seconds
Accuracy of Reading
±1.5% ±0.05 Ω
±1.5% ±0.1 Ω
±2.0% ±0.3 Ω
±3.0% ±1.0 Ω
±5.0% ±5 Ω
±10% ±10 Ω
±20%
Resolution
Accuracy of Reading
±2.0% ±0.05 mA
±2.0% ±0.03 mA
±2.0% ±0.3 mA
±2.0% ±3 mA
Accuracy of Reading
±2.0% ±0.03 A