3) SAFETY INFORMATION
The instrument complies with class II, overvoltage CAT IV 600V, CAT III
1000V of the IEC61010-1 (EN61010-1); IEC61010-2-032 (EN61010-2-032)
standards. Pollution degree 2 for Indoor Use. If the equipment is used in
a manner not specified, the protection provided by the equipment may be
impaired.
EMC: According with EN61326-1: 2006.
BY IEC61010 OVERVOLTAGE CATEGORY INSTALLATION
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equip-
ment supplied from the fixed installation.
Note - Examples include household, office and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed
installation.
Note - Examples includes switches in fixed installations and some equip-
ment for industrial use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at origin of the
installation.
Note - Examples include electricity meters and primary over-current pro-
tection equipment.
In this manual, a Warning identifies conditions and actions that may expose
you to risks or may damage the instrument or equipment testing.
A Note identifies the information that the user should pay particular attention.