5 TROUBLESHOOTING

In the event of faults that are not listed in this chapter, or if the
In the event of faults that ...
suggested procedure does not correct the fault, contact the
PARI GmbH Service Center [See also: 7.4 Contact].
Fault table ...
Fault
The
compressor
does not start
up.
No medication
is coming out
of the
nebuliser.
All LEDs are
flashing.
- 47 -
Possible cause
The power adapter cable
is not plugged into the
device properly.
The power adapter is not
plugged into the mains
socket properly.
The nebuliser nozzle is
blocked.
The connection tubing is
not connected properly.
The connection tubing is
leaking.
There is a temporary
fault in the rechargeable
battery.
Procedure
Check that the power
adapter cable is seated
correctly in the
compressor power cord
socket.
Check that the power
adapter is seated
correctly in the mains
socket.
Clean the nebuliser.
Check that both ends of
the tubing are seated
properly.
Replace the connection
tubing.
Reset the rechargeable
battery as described in
the instructions for use
of the battery.