– Do not operate the compressor in areas where there is a risk
of explosion or in the presence of oxidising gases (e.g.,
oxygen, nitrous oxide, flammable anaesthetics).

3.2 Preparation

Check the compressor housing, the rechargeable battery, the
Check the compressor housing, ...2
12 V car adapter or the power adapter for damage every time
before using. Make sure that the local supply voltage complies
with the voltage data marked on the power adapter identification
label. Only use the compressor if all components are intact. The
compressor must not be operated if the device or the power
adapter or rechargeable battery is damaged, or if damage is
suspected (e.g., after it has fallen or if there is a smell of burning
plastic).
Connecting the nebuliser
• Assemble the nebuliser as described in the instructions for use
of the nebuliser.
• Insert the nebuliser in the holder on
the compressor intended for this
purpose.
• Connect the connection tubing to the
nebuliser and the compressor.
Tip: The tube connectors can be
inserted more easily if they are twisted
slightly when inserting.
- 40 -