Reagent Changes; Adjust Bottle Volumes - Hach 9240 Manual Básico Del Usuario

Ocultar thumbs Ver también para 9240:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 75
Procedure
Verification
Calibration
Electrode reactivation
Temperature
calibration
Refill calibration
solution
Refill reactivation
solution
Refill electrolyte
solution
Electrodes manual
cleaning (with
paper/tissue)
Replace sodium
electrode and
reference electrode
(consumable parts)
Conditioning solution
(non-cationic
applications)
Conditioning solution
(cationic
1
applications)
Solenoid valves
The following items are performed by the Hach Lange service group.
System check-up
External audit

Reagent changes

This procedure is applicable to each change of reagent solution. It allows the regulation of the
volumes of solutions in the bottles. The analyzer consistently checks the consumption of these
solutions, triggering an alarm when these levels get too low. This option must be selected each time
a solution is refilled or changed to reset the measurement parameters.
Note: The conditioning reagent and electrolyte consumption are approximate measurements, so a discrepancy
between measurement and actual consumption may occur. The conditioning solution consumption has been
measured for a pH of 11.2 at an ambient temperature of 25°C.

Adjust bottle volumes

This option should be used to set the default values of reagents or when bottles are being used that
are not full (see
Bottles full
1. Select the ADJUST BOTTLE VOLS. option to set the reagent volumes.
2. Enter the volumes for the conditioning solution, reactivation solution, calibration solution and
electrolyte in milliliters.
1
For additional information, refer to the expanded version of this manual.
24 English
Table 1 Maintenance schedule
Daily
Weekly
Monthly
X
X
X
on page 25).
3
6
Yearly
months
months
X
X
X
X
X
2
As
years
necessary
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X