3. Automated wash
Program the washer using the following parameters:
Phase
Pre Wash
Enzyme Wash
Wash 1
Rinse 1
Dry Phase
Upon completion of the draining phase after Rinse 1, stop the cycle and open the washer
door.
Remove the device from the washer during the thermal phase and place the device back
into the washer for the dry phase.
If necessary, use pressurized air to aid in drying. Visually inspect each device for
cleanliness.
ENZOL® Enzymatic Detergent is validated for cleaning efficacy.
2
Renu-Klenz® is validated for cleaning efficacy.
3
34
Recirculation
Time
2 minutes
2 minutes
2 minutes
2 minutes
7 minutes
Water
Temperature
Cold tap water
Hot tap water
Set point
(66˚C)
Hot tap water
115˚C
Detergent Type and
Concentration
(if applicable)
N/A
Enzymatic Detergent
Non-enzymatic
Detergent
3
N/A
N/A
2