Critical Battery Alarm

The battery has approximately 3 minutes of therapy time remaining. Upon
battery depletion the device will stop delivering therapy and power Off.
Troubleshooting
1. Plug the device into an electrical outlet as soon as possible.
Therapy will continue to be delivered while the device battery is charging.

Battery Failed Alarm

The battery within the device has failed. The device will stop delivering
therapy and power Off. It will only deliver therapy while plugged into an
electrical outlet.
Troubleshooting
1.
If the device has been exposed to temperatures outside its
recommended temperature range, allow device to return to room
temperature.
2.
Plug the device into an electrical outlet; the device will not operate on
battery power. Contact your healthcare provider.
Therapy will continue to be delivered if the device is plugged in.
• Status indicator illuminates yellow.
• Audible alarm sounds every
20 seconds.
• Audible alarm may be paused for
approximately 2 minutes.
• Touchscreen dims to conserve
battery life.
• Status indicator illuminates yellow
when the device is powered On.
• There is no audible alarm.
Note: Battery Failure alarm only
occurs when the device is plugged in.