EN
Canister Full Alarm
The system detects that the canister is full or that the filter inside the
canister is covered with exudate. The device continues to operate but may
not provide the prescribed therapy.
Troubleshooting
1. Pause therapy if active.
2. Replace canister and resume therapy. Refer to "Replacing canister"
section of manual for more details. Continue to next step only if alarm
returns after restarting therapy.
3. Inspect all tubing and connections for any obstructions or kinks.
• If the alarm continues, contact your healthcare provider.
Low Battery Alarm
The battery has approximately 2 hours of therapy time remaining. Upon
battery depletion the device will stop delivering therapy and power Off.
Troubleshooting
1. Plug the device into an electrical outlet as soon as possible.
Therapy will continue to be delivered while the device battery is charging.
• Status indicator illuminates yellow.
• Audible alarm sounds every
20 seconds.
• Audible alarm may be paused for
approximately 2 minutes.
• Status indicator illuminates yellow.
• Audible alarm sounds every
20 seconds.
• Audible alarm may be paused for
approximately 15 minutes.
• Touchscreen dims to conserve
battery life.
32