4.4 How to use the memory function:

• In Clock mode ( Off mode) , press MEM button. The number of stored data
is showed (pic14), sudden the latest result will be displayed with date and
time (pic15).
• Press MEM button repeatedly to review the results measured .
• How to see the average of the last 3 measurements done with T3 System.
(The average is calculated upon last 3 readings measured within 15 minutes).
Please take the measurements following the instructions in the SECTION T3
System of this manual; then, in power off mode, press T3 Button and the
average value of the last 3 measurements made within 15 min. with T3
System, will be displayed with AVG symbol blinking (pic. 16).
NOTE:
The average data using T3 button is shown only if the last 3 measurements have
been done within 15 minutes. Otherwise, the data will be shown as 0.

4.5 To delete all values stored in the memory:

In Clock mode ( Off mode) press MEM button , after the latest data stored will be
displayed press again and hold MEM button for 3 sec. All results will be deleted
display. Press the button MEM or O/I, the display will turn off. (pic.17)
5. ERROR MESSAGE/MALFUNCTION
Follow the below recommended actions and press O/I button again to initiate
another measurement.
Error code
Er 0
Pressure system is unstable
before measurement
Er 1
Fail to detect systolic pressure
Er 2
Fail to detect diastolic pressure
Er 3
Pneumatic system blocked or
cuff is too tight during inflation
Possible cause(s)
Recommended Action
Don't move and try again.
Apply the cuff correctly and try
again
EN
13