Applying Cuff to Patient
Before applying to a patient, read and understand all Warnings in the
Instructions for Use for both your NIBP monitor and cuffs.
1. Select a cuff appropriate for the limb circumference. Size ranges are
printed on each cuff and listed below:
3.1 – 5.7
(cm)
Neonatal 1
Orange
Cuff Text Color
Single-Patient Cuffs
M1866B
2. Place open cuff to arm (or thigh) such that the side with the Philips logo
is away from the patient. Align arrow ( ) labeled Arteria to the
brachial (or femoral) artery (see figure). Ensure the side with the
warning label (
) is toward the patient.
3. Wrap cuff snugly around arm or thigh. Ensure that the index edge falls
within the range indicated by the range arrow, then fasten.
4.3 – 8.0
5.8 – 10.9
7.1 – 13.1
Neonatal 2
Neonatal 3
Neonatal 4
Light Blue
Green
M1868B
M1870B
M1872B
Range Arrow
Boundaries
Index Edge
3
10.0 – 15.0
Infant 5
Navy
Burgundy
M1873B