NOTE: When Legal For Trade is set to On (LFT
Switch in locked position), the menu settings
are affected as follows:
• Calibration (C.A.L) menu is not accessible.
• Dual Scale, Capacity 2, Graduation 2, Filter
and Units are locked at their current settings.
• Zero Range setting is locked at 2%.
• Stable Range setting is locked at 1d.
• Auto-Zero Tracking setting is locked at 0.5d.
• Stable Only is locked On.
• Auto Print/Continuous is disabled.
• Lb:oz is locked at Off.
VERIFICATION AND SEALING
The local weights and measures official or authorized
service agent must perform the verification procedure.
PHYSICAL SEALS
For jurisdictions that use the physical sealing method,
the local weights and measures official or authorized
service agent must apply a security seal to prevent
tampering with the settings. (See Figure 6)
AUDIT TRAIL SEAL
(USA AND CANADA ONLY)
For jurisdictions that use the audit trail sealing method,
the local weights and measures official or authorized
service agent must record the current configuration and
calibration event counter values at the time of sealing.
These values will be compared to values found during a
future inspection.
NOTE: A change to an event counter value is
equivalent to breaking a physical seal.
PAGE 17 OF 69
LEGAL FOR TRADE CONTINUED
The audit trail uses two event counters to record
changes to configuration and calibration settings.
• The configuration event counter (CFG) will index
by 1 when exiting the menu if one or more of the
following menu settings are changed: Dual Scale,
Capacity 2, Graduation 2, Zero Range, Stable
Range, Auto Zero Tracking (AZT), Units (kg, g, oz, lb
or lb:oz) or Stable Printing. Dual Scale, Capacity 2,
Graduation 2 and Zero Range setting is locked
at 2%.
• The calibration event counter (CAL) will index by 1
when exiting the menu if a Span, Lin Calibration or
GEO setting change is made. Note that the counter
only indexes once, even if several settings are
changed.
The event counters can be
viewed by pressing and
holding the MENU button.
While the button is held,
the display will show MENU
followed by Audit.
Release the button when
Audit is displayed to view
the audit trail information.
The audit trail information
is displayed in the format
CFG000 and CAL000.
The scale then returns to
normal operation.
1021 IH-9554