Descargar Imprimir esta página

Ivoclar Vivadent bluephase Manual Del Usuario página 3

Ocultar thumbs Ver también para bluephase:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 7
English
1
PROPER CHARGING
For as long a life cycle of the battery as possible, return the light
to the charging base as often as possible – e.g. after each patient.
2
CLEANING
Battery contacts should be cleaned regularly to ensure that they
are capable of conducting and therefore charging properly – e.g.
by means of mandatory wipe disinfection of the light before each
use.
3
STORAGE / PRACTICE HOLIDAYS
In order to prevent premature wear, always charge the battery
and do not store batteries for longer than 6 months.
4
PROTECT FROM EXTREME TEMPERATURES
For full operational reliability, always use (and store) the battery
at room temperature or between 15 °C (59 °F) and 30 °C (86 °F).
5
REGULAR REPLACEMENT
Since the battery is a consumable, it has to be replaced if the per-
formance drops considerably or after its typical life cycle has
expired after approx. 2½ years.
6
DO NOT PURCHASE BATTERIES IN ADVANCE
In order to avoid irrevocable deep discharge if the battery is not
used, a new battery is best purchased only once the old LiPo bat-
tery needs to be replaced.
7
NO DISPLAY DURING CHARGING
Check if the handpiece has been correctly placed into the charg-
ing base. Clean battery contacts. Information: "Charging Battery"
briefly appears in the display before the display returns to the
stand-by mode.
If cleaning the battery contacts does not improve the results,
remove the battery from the apparatus. Charge the battery sepa-
rately from the handpiece in the charging base for approx.
10 minutes. If the error remains, please contact your dealer or
your local Service Center.
8
SAVETY NOTE
Please note that the lithium polymer battery may react with
explosion, fire, and smoke development in case of inappropriate
handling and mechanical damage.
Please read before use

Publicidad

loading