Troubleshooting Guide - Hoefer HE-PLUS Manual Del Usuario

Tabla de contenido

Publicidad

Troubleshooting Guide
Problem
The LCD screen remains 
blank and the fan does  
not run when the power  
is turned on.
Operation stops with  
alarm: The screen  
displays "NO LOAD".
Operation stops with  
alarm: Display shows  
"OVER VOLTAGE".
Operation stops with  
alarm: Display shows 
"LEAKAGE".
LID alarm.
Cause
AC power cord is not 
connected.
The fuse has blown.
Electrophoresis tank 
is not connected to 
the power supply 
or there is a broken 
circuit in the 
electrophoresis cell.
Buffer concentration 
incorrect.
Circuit is interrupted.
Ground leak detected 
during run.
Cover of gel tank not 
in correct place.
Solution
Check AC power cord connections at both ends. 
Use the correct cords.
Replace the fuse.
Check the connections to the power supply and 
on your electrophoresis cell to make sure the 
connection is intact; check condition of wires 
in electrophoresis unit. Close the circuit by 
reconnecting the cables. Press RUN/PAUSE to 
restart the run.
Replace buffer.
Verify that the running buffer is correct.
Verify the all connections are attached correctly. 
Turn the Power Switch off and on again; restart 
application.
If you cannot restart the instrument, turn off 
the power, disconnect the power cord from the 
outlet, and contact Technical Service.
Check the electrophoresis system for improper 
grounding. Restart the power supply by turning 
the Power switch off and on.
Turn power off then check the gel tank for 
buffer leakage.
Turn off power supply. Place the lid so the 
magnet is pressed against the power supply  
and restart.
If you cannot restart the instrument, turn off 
the power, disconnect the power cord from the 
outlet, and contact Technical Service.