8. WHAT DO I DO IF ...

... no weight is dis-
played during weigh-
ing?
... "0.00" does not
appear before weigh-
ing?
... a segment lights
up continuously or
not at all?
...
is
displayed?
... the display "StOP"
appears?
... the display "bAtt"
appears?
... the display "tEMP"
appears?
... the display "Er" or
"EEEEE"appears
(depending on the
model)?
... the display
"Er:X:11" appears?
... the display
"Er:X:12" appears?
... the display
"Er:X:16" appears?
44 •
The scales are not connected to the power supply.
- Check if the scales are switched on.
- Check if battery block is inserted.
A load was placed on the scales before being switched
on.
- Remove load
- Switch scales off and back on again
The relevant segment is faulty.
Please notify maintenance service.
Voltage of batteries or battery block is dropping.
- Recharge battery block as soon as possible.
- Replace batteries as soon as possible.
The maximum load has been exceeded.
- Remove load.
Batteries or battery block are flat.
- Charge battery block.
- Replace batteries.
The ambient temperature of the scales is too high or too
low.
- Set up scales in an ambient temperature between
+10 °C and +40 °C.
- Wait approx. 15 minutes until scales have adjusted
to ambient temperature.
The automatic software self-test of the scales has found
an error.
- Restart scales.
- If error message appears again, please contact
maintenance service.
Excessive weight has been applied in general or on one
edge.
- Reduce load on scales or distribute weightmore
evenly.
- Restart scales.
The scales have been switched on with too great a load.
- Remove load.
- Restart scales.
Oscillations have been induced in the scales, preventing
determination of the zero point.
- Restart scales.