Technical Data - NUK BP7 Instrucciones De Operación

Ocultar thumbs Ver también para BP7:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 68
Problem
Replacement parts
needed (filter, lip
valve)
The mains adapter included in the
electrical system in accordance with EN 60601-1. All devices in the ME system comply with
the requirements of EN 60601-1 or EN 60950-1.
Medical product according to directive
93/42/EEC
NUK electric Breast Pump
Mains adapter
NUK Teat
NUK Breast Milk Container
Suction
Cycles
Applied parts
Environmental conditions
Atmospheric pressure
operation/storage/transport
Relative humidity:
operation/storage/transport
Temperature
• operation
• storage/transport
Power supply
Inlet
Mains power operation using
adapter
included in NUK packaging type:
PS10J050K2000EU/PS10J050K2000AU/
PS10J050K2000BU
USB power source (not included in NUK
packaging)
Rechargeable battery (Nature Sense model
only)
Battery operating time
Expected service life
Reason

13 Technical Data

NUK packaging
Class I
complies with EN 60601-1, EN 60601-1-2,
CISPR 11 Class 1 Group B
complies with IEC 60950-1
complies with EN 14350
approx. 0 - 350 mbar
approx. 30 - 120 cpm (cycles per minute)
silicone cushion
70 kPa – 106 kPa
15 % – 75 %
+5 °C to +30 °C
+5 °C to +50 °C
USB-C
function, power supply only)
Input: 5 VDC, 2000 mA
mains
Input: 100 - 240 V ~ / 50/60 Hz
Output: 5.0 VDC, 2000 mA
Output 5 VDC, minimum 2000 mA
Recommendation: power bank with at least
8000 mAh
Li-ion, corresponds to IEC 62133,
capacity: 2500 mAh, voltage: 3.7 VDC
1.5 hours
at least 400 h
39
Remedy
Contact your local Service Centre (see
"Contact addresses" on page 193).
and the breast pump form a medical
TM
compatible port (no further USB

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido