Test modes during immunity tests:
Battery mode, standby armed/ready.
Charging mode, actuator not running.
EM immunity standards
Electrostatic discharge
IEC61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields form RF wireless
communication equipment
IEC 61000-4-3
Electrical fast transients / bursts
IEC 61000-4-4
Surges, Line-to-line
IEC 61000-4-5
Conducted disturbance induced by
RF fields
IEC 61000-4-6
Rated power frequency magnetic
fields
IEC 61000-4-8
Voltage dips, short interruptions and
voltage variations
IEC 61000-4-11
Variation of power frequency
IEC 61000-4-28
etac.com
IEC 60601-1-2:2014 test levels
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
IEC 60601-1-2:2014 Table 9
± 1 kV, ±2 kV
100 kHz repetition frequency
± 0,5 kV, ± 1 kV
230 VAC
6 Vrms
0.15 MHz – 80 MHz
80% AM at 1 kHz
30 A/m
50 Hz
0 % U T for 0,5 cycle at 0°, 45°, 90°,
135°, 180°, 225°, 270° and 315°
phase angles
0 % U T for 1 cycle
70 % U T for 25 cycles
0 % U T for 250 cycles
230 VAC
N/A
Additional test levels
± 6 kV contact
20 V/m
800 MHz – 2.5 GHz
80 % AM at 1 kHz
N/A
± 1 kV, ± 2 kV
5 kHz repetition frequency
± 0,5 kV, ± 1 kV
100 VAC and 240 VAC
N/A
30 A/m
50 Hz and 60 Hz
0 % U T for 1 cycle
40 % U T for 10 cycles
70 % U T for 25 cycles
100 VAC and 240 VAC
49 Hz and 51 Hz
100 VAC and 240 VAC, 50 Hz
59 Hz and 61 Hz
100 VAC and 240 VAC, 60 Hz