Medel CHECK Manual Del Usuario página 18

Ocultar thumbs Ver también para CHECK:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 55
Guidance and manufacturer's declaration – electromagnetic emissions
EN
The model Medel Check is intended for use in the electromagnetic environment
specified below. The customer or the user of the model Medel Check should
ensure that it is used in such an environment.
Emission test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Not applicable
IEC 61000-3-3
34
100042_MEDEL CHECK USER MANUAL_ ALL LANGUAGES.indd 34-35
Electromagnetic environment –
guidance
The model Medel Check uses RF
energy only for its internal fun-
ction. Therefore, its RF emissions
are very low and are not likely to
cause any interference in nearby
electronic equipment.
The model Medel Check is suitable
for use in all establishments,
including domestic establishments
and those directly connected to
the public low-voltage power
supply network that supplies buil-
dings used for domestic purposes.
Guidance and manufacturer's declaration – electromagnetic immunity
The model Medel Check is intended for use in the electromagnetic environment
specified below. The customer or the user of the model Medel Check should
ensure that it is used in such an environment.
Immunity
IEC 60601 test level
Compliance level
test
± 6kV contact
± 6kV contact
Electrostatic
discharges
(ESD)
± 8kV air
± 8kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
Not Applicable
transient/
supply lines
burst
.......
IEC 61000-4-4
Surge
± 1kV in differential
Not Applicable
IEC 61000-4-5
mode
Voltage dips,
<5 % UT
Not Applicable
short
(>95 % dip in UT)
interruptions
for 0,5 cycle
and
voltage
40 % UT
variations
(60 % dip in UT)
on power
for 5 cycles
supply
input lines
70 % UT
IEC 61000-
(30 % dip in UT)
4-11
for 25 cycles
<5 % UT
(>95 % dip in UT) for
5 sec.
Power
3 A/m
3 A/m
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage before the application of the test level
EN
Electromagnetic
environment - guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidi-
ty should be at least 30 %.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the model
Medel Check requires conti-
nued operation during power
mains interruptions, it is
recommended that the model
Medel Check be powered
from an uninterruptible
power supply or a battery.
Power frequency magnetic
fields should be at levels cha-
racteristic of a typical location
in a typical commercial or
hospital environment.
35
28/06/2017 10:31:31

Publicidad

Tabla de contenido
loading

Tabla de contenido