9
AUTO REV
10
CAL
11
MEMO
12
SOUND VOLUME Adjusts the sound volume.
13
AC-IN LAMP
14
CHRG LAMP
FISDR-JAP / F1902124.X / 00 / 2011 – created 07/2011
Only for continuous rotary systems, selects one of the 3 auto
reverse modes (see chapter 6.6.4. Auto Reverse Function):
AUTO REVERSING: Automatic stop and reverse motion followed
by forward rotation when preset torque is reached.
AUTO STOP: Automatic stop and reverse motion followed by stop
when the preset torque is reached.
AUTO REVERSE OFF: Auto reverse mode is not activated.
Calibrates the contra-angle to ensure torque accuracy each time
the contra-angle is replaced or lubricated (keep pressed for more
than 2 seconds).
Saves modifications to the speed, torque limit values and auto
reverse mode in each system where modifications are possible
(keep pressed for more than 2 seconds).
Lights green when the control unit is connected to the mains.
Lights or flashes orange while battery is charging or in the refresh
mode (see chapter 6.9. Battery Refresh) and in an error condition
(see chapter 8. Error Code).