Enter the programming and calibration menus by pressing the "CAL SWITCH" and entering the correct
password. (
Liner calibrate.
After you press
display will show LoAd 0, remove all weight from the pan and wait until display show next step. Then
display will show LoAd I, put the weight follow the display, wait for next step.
Follow the display to load calibration weight on the pan.
Until display show LoAd F, load full capacity weight on the pan.
Note: calibration weight regulation: 0<load 1<load 2<load 3<load F(full capacity)
Load1, Load2, Load3 need be integer weight.
In TECH always press the push button and will be show the pin.
Version 1.16:
Press
Press
After stable and zero indicator on, display will be show LOAD 1.
Place the first calibration mass weight on the pan.
After stable and zero indicator on, display will be show LOAD 2.
Add again the second calibration mass weight on the pan.
After stable and zero indicator on, display will be show LOAD 3.
Add again the third calibration mass weight on the pan.
After stable and zero indicator on, display will be show LOAD 4.
Add again the fourth calibration mass weight on the pan.
P1 Lin
After stable and zero indicator on, display will be show LOAD 5.
Add again the fifth calibration mass weight on the pan.
After stable and zero indicator on, display will be show LOAD 6.
Add again the sixth calibration mass weight on the pan. (*Note: Calibration mass weight
should be full capacity of the scale)
After stable and zero indicator on, display will be show LOAD 0. Remove the calibration mass
weight from the pan. (*Note: Ensure the pan is empty)
After stable and zero indicator on, display will be show LOAD 6.
Place the calibration mass weight on the pan (*Note: Calibration mass weight should be full
capacity of the scale)
After stable and zero indicator on, display will be show LOAD 5.
Remove the sixth calibration mass weight on the pan.
After stable and zero indicator on, display will be show LOAD 4.
Remove the fifth calibration mass weight on the pan.
After stable and zero indicator on, display will be show LOAD 3.
Remove the fourth calibration mass weight on the pan.
After stable and zero indicator on, display will be show LOAD 2.
Remove the third calibration mass weight on the pan.
After stable and zero indicator on, display will be show LOAD 1.
Remove the second calibration mass weight on the pan.
After stable and zero indicator on, display will be show LOAD 0.
Remove the first calibration mass weight on the pan. (*Note: Ensure the pan is empty)
After stable and zero indicator on, display will be started self checking. After checking display will
come to normal weighing mode.
Note: During the linear calibration steps LOAD 1 to LOAD 6, mass weights want to add on the
pan. Don't remove all mass weights for the next step.
During the linear calibration steps LOAD 6 to LOAD 1, mass weights want to reduce from the
pan. Don't remove all mass weights for the next step.
36
,
and
) and press
key, display will show Pin again, press
key to enter calibration, display will be shown PIN
keys, display will be show LOAD 0 (*Note: Ensure the pan is empty)
,
and
key to enter
,
,
USER MANUAL BAR
key and
key to enter,
EN
EN