Appendix; Error Code; Weighing Unit Listing; Trouble Self Checking - Baxtran BAR Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

9. APPENDIX

9.1. ERROR CODES
During the initial power-on testing or during operation it is possible the scale may show an error message. The meaning of the
error messages is described below.
If an error message is shown repeat the procedure that caused the message, turning the balance on, calibration or other functions.
If the error message still is shown then contact your dealer for further support.

ERROR CODE

9.2 WEIGHING UNIT LISTING

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15

10. TROUBLE SELF CHECKING

PROBLEM
Display is blank, No turn on test
38
DESCRIPTION
Err 3
Calibrate weight error
Err 4
Initial Zero is greater than allowed (4% of
maximum capacity) when power is turned
on or when the
Err 6
A/D count is not correct when turning the
scale on
Err 7
Percentage setting error
Err 8
External calibration weight error
UNIT SYMBOL
g
g
ct
lb
oz
d
ozt
dwt
mom
tL.H
tl.c
tl.t
t
bt
n
POSIBLE CAUSE
Mains power is turned off.
Power supply not plugged in.
Power supply faulty.
Internal battery not charged.
Display turned off.
Use incorrect calibrate weight
Weight on the pan when turning the scale
on.
key is pressed.
Excessive weight on the pan when zeroing
the scale.
Improper calibration of the scale. Damaged
load cell.
Damaged electronics.
Platform not installed.
Load cell damaged.
Electronics damaged.
0.01% weight must>0.5d
More than 4% error as linear calibration
NAME OF THE UNIT
gram
gram (high resolution)
carat net
pound
ounce
drachm
ounce troy
scruples
momme
tael twn
tael chn
troy tael
tola
bangladeshi tola
newton
USER MANUAL BAR
POSSIBLES CAUSES
CONVERSION
= 0,2 g
= 453,59237 g
= 28, 349523000 g
= 1,7718450 g
= 31,1034768 g
= 1,55517400 g
= 3,749996 g
= 37,49996 g
= 37,799375 g
= 37,4290018 g
= 11,6638039 g
1bt= 11,6638039 g
1bt= 16Ana= 96Roti
1n= 102,04 g
COMMON SOLUTIONS
Check power is going to the scale and
switch is on.
Verify the voltage going to the scale
matches the power supply labels on
the power module or scale.
EN
EN

Publicidad

Tabla de contenido
loading