Use - Agilent Technologies TV 303 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 49

Use

This paragraph details the fundamental operating procedures.
Make all electrical an pneumatic connections before the use of the
system.
While heating the vacuum chamber, the temperature of the inlet
flange must not exceed 120 °C.
WARNING!
Never use the turbopump when the inlet flange is not connected to the
vacuum chamber or is not blanked.
Do not touch the turbopump or any of its accessories during the heating
process. The high temperatures may cause burns.
Avoid impacts, oscillations or harsh movements of the pump when in operation.
CAUTION!
The bearings may become damaged. Use air or inert gas free from dust or
particles for venting the pump. The pressure at the vent port must be less than 2
bar (above atmospheric pressure). For pumping aggressive gases, these pumps
are fitted with a special port to allow a steady flow of inert gas (like N2, Ar) for
pump bearing protection (see the appendix "Technical Information").
WARNING!
When employing the pump for pumping toxic, flammable, or radioactive
gases, please follow the required procedures for each gas disposal.
Do not use the pump in presence of explosive gases.
TV 303 User Manual / 87-901-019-01
(A)
Instructions for Use

Use

203/268
16

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

969-8964969-8998

Tabla de contenido