Alarm Indicators and Screen Symbols
Note: All SimplyGo alarms are low priority alarms� The alarm system should
be verified before use and between users by service personnel in
accordance with the SimplyGo service manual�
Technical Alarms
Visual, Audio
Indicators
A sequence of
3 short
beeps
+
1 long beep
that repeats every
4 seconds
22
Description
No Breath Alarm
This alarm occurs
when a breath is not
detected for a period
of 2 minutes or more�
It can be silenced by
pressing the Alarm
symbol� This alarm
resets itself as soon as
a breath is detected� If
no breath is detected
after approximately
30 minutes, the device
shuts down to conserve
power�
Note: This alarm is
inactive for ten minutes
after startup� Upon
startup, this alarm could
take a maximum of 12
minutes to activate
from the time the alarm
condition is present�
Otherwise, it can take
up to 2 minutes to
activate�
What To Do
Check the connection
from the cannula to
the device�
Make sure the nasal
cannula is properly
positioned on your
face and that you are
breathing through
your nose�
Make sure the
cannula tubing is not
kinked or obstructed�