Disinfection
The exterior covers of the device and the handle of the rechargeable battery
should be disinfected weekly and between patient use as follows:
1�
Clean the device as indicated above�
2�
To disinfect, use a household chlorine bleach containing 8�25% sodium
hypochlorite� Combine 9 parts water to 1 part bleach�
3�
Using a damp cloth with the bleach solution, wipe the exterior surfaces�
4�
Allow the surface to remain damp for 2 minutes� Wipe dry as necessary�

Accessories Cleaning and Maintenance

Humidifier and Cannula
Clean and replace the humidifier and cannula as instructed by the
manufacturer and your equipment provider�

Carrying Case

The carrying case should be cleaned monthly by performing the following
steps:
1�
Remove the SimplyGo device from the carrying case�
2�
Use a dry, soft-bristled brush gently to remove any visible soil on the
carrying case before wiping�
3�
Use a cloth dampened (not dripping) with a mild, household liquid
dishwashing solution to clean the exterior of the carrying case� Ensure
you remove all visible soil; a damp, soft-bristled brush may be used if
necessary�
4�
Perform a 'rinse' using a cloth dampened (not dripping) with warm water
to wipe the exterior of the carrying case�
5�
Allow the case to air dry�
The carrying case should be single use and replaced between patients�

Device Storage

S tore your device in a place where it will remain clean and dry�
Caution: Do not store your device or accessories in extreme temperatures,
below -4° F (-20° C) or above 140° F (60° C)�