Conductivity Calibration
Type = 1 point
h
1.25
H
µS/cm
25.0
A
°C
A Point1 = 0.055 µS/cm
A
C = 0.057 µS/cm
1.25
H
µS/cm
25.0
A
°C
C M=0.1000
A=0.0000
Save Adjust
24
Select 1 POINT instead of 2 POINT CALIBRATION by pressing the . key once followed
by the [ENTER] key.
u
CAUTION: Rinse sensors with a high-purity water solution before every calib ration to
prevent contamination of the reference solutions.
Place the electrode in the reference solution.
Enter the value of Point 1 and press the [ENTER] key.
u
After the calibration of the cell multiplier or slope calibration factor "M" i.e. cell con-
stant and the Adder or offset calibration factor "A" are displayed.
For ISM (digital) sensors select ADJUST, CALIBRATE or ABORT to finish calibration.
For Analog sensors select ADJUST or ABORT to finish calibration. See Chapter 6
u
"General Setup" on Page 13.