pH Calibration

ISM (digital) sensors
Analog sensors
(PATH: Cal)
After every successful calibration the following options are available.
After selection one of these option the message "RE-INSTALL SENSOR and Press
[ENTER]" appears on the display.
After pressing [ENTER] the M400 returns to the measuring mode.
Adjust
Calibration values are stored in the sensor and used for the measurement. Addition-
ally, the calibration values are stored in the calibration history.
Calibrate
Calibration values are stored in the calibration history for documentation, but not be
used for the measurement. The calibration values from the last valid adjustment are
further used for the measurement.
Abort
Calibration values are discarded.
Adjust
Calibration values are stored in the transmitter and used for the measurement. Ad-
ditionally, the calibration values are stored in the calibration data.
Calibrate
The function "Calibrate" is not applicable for analog sensors.
Abort
Calibration values are discarded.
15