5
General Setup (applies for all parameters)
21.7
%sat
25.0
°C
MENU
Quick Setup
Channel configuration (for multi-channel devices)
A
A
6.0
pH
25.0
°C
A Sensor Type = pH/ORP
B Sensor Type = _
A
A
6.0
pH
25.0
°C
a =
pH
Analog Output
Yes
Analog out puts
7.0
pH
25.0
°C
Aout1 min= 0.000 pH
ì
Aout1 max= 14.00 pH
While in Measurement mode press the [MENU] key to bring up the Menu selection.
Select "Quick Setup" and press the [ENTER] key.
Display Convention:
1st line on display
2nd line on display
Select the units of measurement for a and b. Only lines a and b can be configured in
Quick setup. Go to the Configuration Menu to configure lines c and d.
Select the sensor type for input channel A (Connector TB3) and input channel B
(Connector TB4). You can configure any mixture of parameters for the two channels.
The selection has to reflect the connection of sensors to the M300.
The user can configure the second channel by changing a to c, using the
Select the units of measurement for a and c. Only lines a and c can be configured in
Quick setup. Go to the Configuration Menu to configure lines b and d.
By using analogue 2-electrode or 4-electrode conductivity sensors,
see "Conductivity Quick Setup" below for intermediate steps.
By selecting Yes the linear 4–20 mA analog output Aout1 will be set up for measure-
ment a when [ENTER] is pressed. Selecting No means that no analog output is set up.
Aout1 min, Aout1 max are the minimum and maximum measurement values for the
4 and 20 mA values respectively.
Note for multi channel devices: The user can configure output A out 3 to
mesurement c by going back to the previous menu and selecting c.
(PATH: Menu/Quick Setup)
a
3rd line on display
b
4th line on display
c
d
or q key.
9