Conductivity Quick Setup

For Display setup, analog outputs and setpoints see the section "General Setup".
Channel Selection
A
A
1.25
MΩ-cm
25.0
°C
Calibrate Sensor
Channel B Conductivity
Sensor Type Selection
1.25
µS/cm
25.0
°C
Sensor Type = Cond(2)
Cell Constant
1.25
µS/cm
25.0
°C
p M=0.1003
A=0.0000
s M=1.0000
A=0.0000
Measurement units
1.25
µS/cm
25.0
°C
a S/cm
Analog Output?   Yes
14
(PATH: Menu/Quick Setup)
For multi-channel devices: Using the
user change the channel to be calibrated.
Select the type of sensor to be used with the M300 transmitter. Choices are "Cond(2)",
used for all 2-Electrode type sensors and "Cond (4)" for all 4-electrode sensors. Press
the [ENTER] key.
Enter the appropriate cell constant(s):from the sensor label or certificate (M) for 2-elec-
trode sensors, leaving (A) at 0.000; or (M) and (A) values for 4-electrode sensors.
Press the [ENTER] key.
Select the measurement (conductivity or temperature) and units for measurement.
If using analog output, select Yes. Refer back to section "General Setup" to continue
setup.
or q key on the "Channel A" field lets the