6
General Setup
Configuration tool or asset-management tool
h
10
You configure the M100 transmitter either via a configuration tool, an asset-manage-
ment-tool or via a HART handheld terminal.
Prerequisite: The M100 transmitter and the sensor are mounted and electrically con-
nected.
Note: The DTM and the configuration tool PACTWare™ are on the supplied CD-ROM.
You can also download the DTM via Internet "www.mt.com/M100".
For steps 1 to 5 refer to the documentation about the tool.
1.
Install configuration tool e.g. PACTWare™ or asset-management-tool.
2.
Install DTM for HART interface and DTM for M100 transmitter.
3.
Update device catalog.
4.
Build up connection. Check COM port settings if necessary.
Set Tag or/and Long tag.
5.
Menu path: Device Setup > Detailed Setup > HART Info
6.
Set Date and Time. Set the time in the 24 hour format. The time format can-
not be changed. Menu path: Device Setup > Detailed Setup > System
7.
Set range of the analog output signal.
Menu path: Detailed setup > Output Condition > Analog output > Range
URV (Upper range value) and LRV (Lower range value), can be changed
–
to the current measurement range
–
USV (Upper range sensor) and LSV (Lower range sensor), defined by the
sensor and cannot be changed
Define the process variables PV, SV, TV and QV
8.
Menu path: Device Setup > Detailed Setup > Measurements > Channel Setup