TensCare Sonic Stim + Instrucciones De Uso página 17

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

6.2.2 Combination therapy
1)
Connect the lead wire and
electrode pad to the unit
as shown by the illustrations
to the right.
Plug the lead wire into
the connection point
attached
device,
Connect the electrode pad with the lead wire.
Make sure all connections are securely in place.
Caution: The device must be turned off before connecting the lead
wires to the device.
2)
Place electrode firmly on the skin after cleaning
and drying the treatment area.
Place the electrode pad on the area of the
body indicated by your physician or therapist.
Make sure the electrode pad is placed firmly
to the skin and has made good contact
between the skin and the pad.
3)
Apply Transmission Gel
Apply a layer of ultrasound transmission gel on the
treatment area. The gel acts as a coupling substance
and ensures treatment effectiveness. The area treated
should be two times the diameter of the treatment
head
Caution: Never apply the gel to the applicator. The applicator will
register this as contact and may emit ultrasound energy, which
could damage the applicator.
4)
Switch on the device
Connect the power adapter according to section 5.3.
Switch on the device, using power ON/OFF switch by
switching from OFF to ON position. The Power LED will
light green.
to
the
15

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido