5)
Adjust ultrasound intensity
Press MODE button to select ultrasound intensity within
low, middle and high. The LED will be lighted to indicate
the selected intensity.
6)
Adjust treatment time
Press the TIME button to cycle through the treatment
time (5, 10 and 15 minutes), as shown by the Time LED
indicators. When the time is chosen, the system will start
working. During working time, the user can press the
TIME button to adjust the treatment time.
7)
Adjust stimulation intensity
Press the + button to increase the intensity of the
stimulation. Press the - button to decrease the intensity
of the stimulation. The STIM LED will flash every time the
+ or - button is pressed.
Note: The colour of the LED indicates the output
stimulation intensity.
Green light: Output intensity < 10 V;
Orange light: Output intensity :> 10 V.
Caution: The electrical stimulation cannot work without the
ultrasound output. Users can only adjust the stimulation intensity
after the ultrasound is working properly.
Caution: If the stimulation level is uncomfortable or becomes
uncomfortable, reduce the stimulation intensity to a comfortable
level and contact your medical practitioner if problems persist.
Caution: Move the ultrasound treatment head while you are
adjusting the stimulation intensity to prevent the local skin
temperature from becoming too high or burning.
Caution: Each step of increase output intensity is 1 V when the
output intensity less than 5V; 0.5V/step when the output intensity is
over 5V.
Caution: In the particular standard for Electrical Nerve and Muscle
Stimulators, IEC 60601-2-10, it is recommended not to exceed a
16