Installation
5.4.2
Installing a Device with Round Draft
Shield
Procedure
Place the round glass ring on the base of the draft
t
shield.
Place the cover of the draft shield onto the round
t
glass ring.
5.5
Acclimatizing the Device
When a cold device is brought into a warm environment:
The temperature difference can lead to condensation
from humidity in the device (moisture formation).
Moisture in the device can lead to malfunctions.
Procedure
Allow the device to acclimatize for approx. 2 hours at
t
the installation site. Ensure that the device is
disconnected from the power supply during that
time.
18
Entris
BCE Operating Instructions
®
Place the base of the
t
draft shield into the
round recess on the
device.
Fit the pan support
t
on the pan retainer.
Place the weighing
t
pan onto the pan
support.