Parameter
Setting values
CAL.-ADJ.
ON Z/T
ON*
OFF
ISOCAL
OFF
NOTE
ON*
EXT.CAL.
UNLCKED*
LOCKED
* Factory setting
7.3.2

"SETUP" / "GEN.SERV." Menu

Parameter
Setting values
MENU.RES.
DEFAULT
NO*
* Factory setting
7.3.3

"DEVICE" / "RS232" Menu

Parameter
Setting values
DAT.REC.
XBPI
SBI*
REM.DISP.
BARCODE
YDP20
YDP30
OFF
BAUD
600, 1200, 2400,
4800, 9600*,
19200, 38400,
57600, 115200
* Factory setting
Explanation
The adjustment must be started or exited manually after calibration with the
[Confirm] button.
Activates the initial taring / zeroing. The device is tared or zeroed after it is
switched on.
Deactivates the initial taring / zeroing: After it is switched on, the device
shows the same value as before it was last switched off.
Switches the isoCAL function off.
If the balance needs to be calibrated: The [isoCAL] button flashes. The iso-
CAL function must be manually triggered with the [Adjust] button.
Activates the isoCAL function. The device is automatically adjusted as soon
as a trigger starts the isoCAL function.
Activates the external calibration / adjustment function under [CAL./ADJ.].
Deactivates the external calibration / adjustment function under [CAL./
ADJ.].
Explanation
Resets the system settings to the factory default settings.
Deactivates the option of resetting the device menu.
Explanations
Extended range of commands to control numerous balance functions with
binary protocol for direct communication with the device.
Enables SBI communication. The data is output to a PC or control unit.
Enables the use of ESC commands from a PC to control the basic balance
functions with ASCII protocol.
Enables data output on another display.
Allows for the connection of an approved barcode scanner.
Sets the standard settings of YDP20 printers.
Sets the standard settings of YDP30 printers.
Deactivates the automatic data output.
Sets the baud rate to the selected value.